Scope: cannot connect, or shows "Connected" but settings are not applied
Set the resource name to VICP::<scope IP>::INSTR (IP: Internal information). On the scope, check the control method (TCPIP (VICP) or LXI (VXI11)) and IP under Utilities → Utilities Setup → Remote. If ping works, VICP should connect. On connect, the *IDN? response is verified; if there is no response, VICP ⇔ VXI-11 is tried automatically before reporting an error
Stage: ping works but the connection times out
The TCP port number does not match the controller setting. Enter <stage IP>:<port> in the address field, using the values in Internal information. If a previous session is holding the connection, restart the controller. If it connects but does not respond, check the SHOT-compatible mode and the delimiter (CRLF)
Stage: ping fails after a network change (the scope responds)
Check that the control PC's subnet mask is 255.255.255.0 (Initial device settings). Check in the controller settings that the SHRC-203's subnet mask and gateway put it on the same network as the PC, and power-cycle the controller after changing its settings (they do not take effect until restart). To check that no other device has the same IP, look at arp -a in a terminal on the PC. Addresses are in Internal information
Laser: cannot connect
Check the COM number in Device Manager (COM5 on the control PC). Check that the FTDI VCP driver is installed
"Machine coordinates: unknown" is shown and you cannot go to the saved origin
The GUI lost track of the machine coordinates (M-ORG from the JS-301, controller restart, interrupted homing, etc.). Run [Mechanical Home (H:W)] from the GUI
Origin finding stops with "No clear dip" or "No clear edge"
Check that the beam is on the reference strip, 50–200 µm inside its end, that the laser is ON, and that a signal is visible; then run again
"Valley too shallow", or the Z valley cannot be found
Widen the Z range to ±30–50 µm, or roughly focus with the camera first, then run again
The Z residual of the full calibration exceeds a few µm, or the coefficients changed a lot since last time
Run it again, or use the "Measured Z" mode for Z tracking in the scan
After origin finding or full calibration the right column becomes wider, or the tilt coefficient display is stale
Bug in an older version (fixed in the 2026-10 revision). Use the latest version
The jog direction does not match the camera image
The camera or stage mounting orientation was changed. Fix stage_screen_dir in config_local.json
The scan stops automatically if the number of files on the scope does not increase for 2 points in a row. Check the scope's Save Waveform setup and the save folder
timeout lines in the scan log
Acquisition did not finish in time. Check that the laser is ON and that the EXT trigger is firing. Also check that [Averaging OFF] has not been unchecked
Warning at start that "Z is 30 µm or more away"
Check that this is intended. Origin finding may not have been done, or the Z tracking mode may be wrong
Adjust [Gain] on the camera panel (the communication log shows the actual value, so you can see whether it took effect; if the actual value does not change, the camera does not support it). If that does not help, adjust the illumination brightness and the lens aperture