Standard settings

Standard measurement conditions and GUI defaults. The values come from config.py (standard conditions fixed by measurements in 2026-07) and the GUI source code.

Sensor and bias

Item Value Notes
HV (sensor bias) −190 to −195 V (at a sensor temperature of 33 °C). Changed with temperature. Limit −220 V Matsusada P4L650-0.1, manual. The reference value in the GUI status bar is −185 V
LV (readout board) 6.0 V (fixed) Rohde & Schwarz NGP800 series, manual. CH1
Current limits HV 1.0 mA, LV 1.0 A Normal currents: HV about 0.4–5 µA (display 0.0 mA), LV about 0.450 A
Power-on order After the other equipment, LV → HV The other equipment can be powered on in any order
Readout C1=J8, C2=J7, C3=J6, C4=J5 Standard sensor

Oscilloscope (contents of [★ Apply Standard Settings ★])

Item Value
Trigger Source EXT, coupling DC, slope POS, level 410 mV
Timebase 1 ns/div
Horizontal offset −53 ns (display range +48 to +58 ns)
Vertical C1–C4: 50 mV/div, offset +100 mV (display range −300 to +100 mV)
Input coupling DC 50 Ω (D50)
Averaging OFF
Sampling 10 GS/s (setting in the analysis note)

Laser

Item Value Notes
Trigger source internal (ts=0) The laser is the reference clock
Rep. rate 1 MHz The GUI standard since the 2026-11-01 version (the preset reads "1 MHz (standard)"). In older versions the initial value was 1.25 MHz
TUNE manual 50 % Adjusted to MIP-equivalent at 1 MHz. 0 % strongest, 100 % weakest
One-click setting [★ Apply Standard Settings (1 MHz / TUNE manual 50% = MIP equiv.) ★] Sends the three items above (internal, 1 MHz, manual 50 %) in one go. The values come from laser_freq_hz, laser_tune_auto and laser_tune_percent in config.py
Camera light OFF during measurement Operated by hand inside the dark room
TUNE optimization: single-CH target 200 mV (PKPK)
TUNE optimization: adjacent 2-ch sum target 64 mV MIP-equivalent value at the strip midpoint from the DESY beam test (2026)
TUNE optimization: tolerance / coarse sweep step ±1 mV / 2 % Fine sweep in 0.5 % steps

Stage

Item Value
Travel 2 µm/pulse (X, Y and Z, confirmed)
Speed Start 1000 pps, max 10000 pps (= 20 mm/s), acceleration 200 ms
Jog steps 2, 4, 10, 20, 50, 100, 500, 1000, 5000 µm

Alignment

Item Value
Z Focus Range ±20 µm, step 2 µm, mode "Min amplitude (focus)"
Z measurement position On silicon, 80 µm beside the electrode center
X-edge detection position 50 µm beside the electrode center
X design offset (signal edge → frame edge) 632 µm
Fine finishing 1-pulse (2 µm) steps, 4 reads per point (median), up to 2 verifications
Full calibration points (standard sensor) X = −5.3, −9.9 mm, Y = C2, C4

Tilt coefficients (defaults for the standard sensor)

Coefficient Value
dZ/dX −0.02 µm/mm
dZ/dY +3.79 µm/mm
dY/dX +7.67 µm/mm

Auto scan

Item Default
Fast axis Y +750 → −2250 µm, 10 µm steps
Slow axis X −632 → −9999 µm, 500 µm steps
Position step for 1D + TUNE 50 µm (set automatically when the mode is selected)
TUNE sweep 50 → 90 %, 5 % steps
Sequence 2000 segments/point
Settle / Timeout 0.5 s / 60 s
Z tracking Coefficients (focus at origin)
Rot. tracking (Y) ON
Snake (bidirectional) ON (2D)
prefix scan
Laser ON at start / Laser OFF at end OFF / ON

Network and connections

Device Address
Control PC, oscilloscope, stage, NAS LAN. IP addresses and user names: Internal information (password protected)
Oscilloscope VICP
Laser PiLAS DX COM5 (19200 baud, 8N1, CR terminated)
Camera Device index 0