Standard settings
Standard measurement conditions and GUI defaults. The values come from config.py (standard conditions fixed by measurements in 2026-07) and the GUI source code.
Sensor and bias
Item
Value
Notes
HV (sensor bias)
−190 to −195 V (at a sensor temperature of 33 °C). Changed with temperature. Limit −220 V
Matsusada P4L650-0.1, manual. The reference value in the GUI status bar is −185 V
LV (readout board)
6.0 V (fixed)
Rohde & Schwarz NGP800 series, manual. CH1
Current limits
HV 1.0 mA, LV 1.0 A
Normal currents: HV about 0.4–5 µA (display 0.0 mA), LV about 0.450 A
Power-on order
After the other equipment, LV → HV
The other equipment can be powered on in any order
Readout
C1=J8, C2=J7, C3=J6, C4=J5
Standard sensor
Oscilloscope (contents of [★ Apply Standard Settings ★])
Item
Value
Trigger
Source EXT, coupling DC, slope POS, level 410 mV
Timebase
1 ns/div
Horizontal offset
−53 ns (display range +48 to +58 ns)
Vertical
C1–C4: 50 mV/div, offset +100 mV (display range −300 to +100 mV)
Input coupling
DC 50 Ω (D50)
Averaging
OFF
Sampling
10 GS/s (setting in the analysis note)
Laser
Item
Value
Notes
Trigger source
internal (ts=0)
The laser is the reference clock
Rep. rate
1 MHz
The GUI standard since the 2026-11-01 version (the preset reads "1 MHz (standard)"). In older versions the initial value was 1.25 MHz
TUNE
manual 50 %
Adjusted to MIP-equivalent at 1 MHz. 0 % strongest, 100 % weakest
One-click setting
[★ Apply Standard Settings (1 MHz / TUNE manual 50% = MIP equiv.) ★]
Sends the three items above (internal, 1 MHz, manual 50 %) in one go. The values come from laser_freq_hz, laser_tune_auto and laser_tune_percent in config.py
Camera light
OFF during measurement
Operated by hand inside the dark room
TUNE optimization: single-CH target
200 mV (PKPK)
TUNE optimization: adjacent 2-ch sum target
64 mV
MIP-equivalent value at the strip midpoint from the DESY beam test (2026)
TUNE optimization: tolerance / coarse sweep step
±1 mV / 2 %
Fine sweep in 0.5 % steps
Stage
Item
Value
Travel
2 µm/pulse (X, Y and Z, confirmed)
Speed
Start 1000 pps, max 10000 pps (= 20 mm/s), acceleration 200 ms
Jog steps
2, 4, 10, 20, 50, 100, 500, 1000, 5000 µm
Alignment
Item
Value
Z Focus
Range ±20 µm, step 2 µm, mode "Min amplitude (focus)"
Z measurement position
On silicon, 80 µm beside the electrode center
X-edge detection position
50 µm beside the electrode center
X design offset (signal edge → frame edge)
632 µm
Fine finishing
1-pulse (2 µm) steps, 4 reads per point (median), up to 2 verifications
Full calibration points (standard sensor)
X = −5.3, −9.9 mm, Y = C2, C4
Tilt coefficients (defaults for the standard sensor)
Coefficient
Value
dZ/dX
−0.02 µm/mm
dZ/dY
+3.79 µm/mm
dY/dX
+7.67 µm/mm
Auto scan
Item
Default
Fast axis Y
+750 → −2250 µm, 10 µm steps
Slow axis X
−632 → −9999 µm, 500 µm steps
Position step for 1D + TUNE
50 µm (set automatically when the mode is selected)
TUNE sweep
50 → 90 %, 5 % steps
Sequence
2000 segments/point
Settle / Timeout
0.5 s / 60 s
Z tracking
Coefficients (focus at origin)
Rot. tracking (Y)
ON
Snake (bidirectional)
ON (2D)
prefix
scan
Laser ON at start / Laser OFF at end
OFF / ON
Network and connections
Device
Address
Control PC, oscilloscope, stage, NAS
LAN. IP addresses and user names: Internal information (password protected)
Oscilloscope
VICP
Laser PiLAS DX
COM5 (19200 baud, 8N1, CR terminated)
Camera
Device index 0